Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A016 - Rotation Movement Troubleshooting
Page no. 1065
Chapter 9
JC-TSG-A-016.fm
3-4-3
Locate dry surfaces
•
Rotate the examination arm between -165°
and +185°.
Check for dry or partly dry surfaces as shown
here (4); this is the visible effect of incorrect
greasing
! Notice:
Be sure to identify dry or partly dry surfaces before greasing. General greasing may not help! The
most critical areas are visible with the arm in the following positions: 0°, 45°, 135°, and 180°.
4