Chapter 9
Page no. 1068
JC-TSG-A-016.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A016 - Rotation Movement Troubleshooting
3-7
Rotation Motor Checks
Check cabling connections and connectors on the Rotation Board, and the Rotation Motor.
•
If the Rotation Motor connectors are not OK, change the Rotation Motor according to
Note:
To confirm whether the Rotation Motor is faulty you can additionally measure the resistance be-
tween the motor connector pins (J10 on the Rotation Board). If the resistance is infinite (open cir-
cuit), the Rotation Motor is faulty.
•
If the Rotation Board connectors are not OK, change the Rotation Board according to
.
•
If the cabling connections and connectors are OK but the Rotation does not still work, change the
Rotation Board according to
Job Card D/R A221 - Rotation Board
, since it is the last
remaining part that can be replaced.