Chapter 9
Page no. 1086
JC-TSG-M-028.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG M028 - E24 084/008 Error Troubleshooting
ILLUSTRATION 1 - E24 FAILURE BEFORE EXPOSURE 084/008 TROUBLESHOOTING
Intermittent
E24 084/008
At AOP pre-
exposure
Check error log
You may confirm that
you have the “tube arc-
ing” issue by checking
the tube spit number
Pre-exposure stopped by tube arcing
1. Perform
Job Card TSG M022 - HV Troubleshooting on page 1403
2. See
,
Preventive Actions that May Reduce Tube Arcing
.3. Write down the current acquisition number (See
).
Bucky to Main Pos CPU Synchronization Error
Check/Replace Positioner CAN bus
1 W311 Distribution Board PL301 J3a to Bucky Board PL302 J1A
2 W314 Distribution Board PL301 J1 - to Interf. Board PL104 J5
POS to DMR Synchronization Error
1. Check/Replace optical fiber be-
tween Generator and Gantry W121
PL104-J8 to 400PL2-XJ9
2. Check/Replace Gantry interface
board PL104
3. Check/Replace Generator interface
board 400PL2
Solve the
issue?
A
Only
with
Bucky?
POS 170/007 Bucky init error
or
DMR 027/003
at +/- 2s?
X-Ray started
(X-Ray end: beep heard)?
mAs value displayed
Change Bucky (see
Yes
Yes
Yes
Yes
No
No
No
No
Do not know
Do not know
B
Could be
B
before
A
if you can check
another Bucky.
No
System Observation:
Observe system during at least
one month (See