GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A229 - Rotation Brake
Page no. 1247
Chapter 9
JC-DR-A-229.fm
Job Card D/R A229 - Rotation Brake
Chapter 9
1
SUPPLIES
•
•
Two transportation screws (found either in your service kit or supplied with the new brake).
M5x50 ISO 4762 - total length 50 mm, length of thread 22 mm.
2
TOOLS
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 90 minutes
4
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5
PREREQUISITES
•
Switch off the Gantry power.
•
Remove the top right and left column covers and
remove the front covers (refer to
- Remove/Reinstall Gantry Covers
).
•
Turn the Gantry power on.
•
Raise the arm to have good access to the Rotation
Brake (1) from the rear of the Gantry.
•
Switch off the Gantry power.
•
Loosen and remove the Gantry curtain (refer to
Card D/R A208 - Gantry Curtains
).
1