Chapter 9
Page no. 1366
JC-DR-A-266.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A266 - Tube Tilt Board
6
PROCEDURE
6-1
Disassembly of the Tilt Stepper Board
1. Unplug all cables from the Tilt Board, after noting
their positions.
2. Remove the four screws (2) securing the Tilt Board.
(2.5 mm allen wrench).
3. Remove the Tilt Board.
6-2
Reassembly of the Tilt Stepper Board
Reverse the disassembly procedure.
2