Job Card D/R A411 - Mains Distribution Board 200-PL1
Page no. 1605
Chapter 9
JC-DR-A-411.fm
6-2
Installation of the Mains Distribution Board 200-PL1
1. Carefully place the new Mains Distribution board on the Generator chassis. Ensure that the holes for
the four bolts line up with the four threads on the Generator chassis.
2. Attach the four bolts (1) (7 mm allen wrench) to secure the Mains Distribution board to the Generator
chassis.
3. Connect all the cables to the new Mains Distribution board as summarized in the diagram and table
below.
Connec-
tor Label
Tool Description
of
cable
(usually identified by code on a white label)
Cable
color
XJ1
8 mm open ended wrench
XJ1
Brown
XJ2
8 mm open ended wrench
Two cables: XJ2
Brown
XJ3
8 mm open ended wrench
XJ3
Brown
XJ4
8 mm open ended wrench
XJ4
Brown
XJ6
8 mm open ended wrench
XJ6
Brown
XJ7
8 mm open ended wrench
XJ7
Brown
XJ8
8 mm open ended wrench
XJ8
Brown
XJ9
8 mm open ended wrench
XJ9
Brown
XJ10
N/A
XJ10/PL1 2111145
White
XJ11
N/A
200 PL1 XJ1
Gray
XJ12
N/A
Empty
N/A
XJ13
N/A
Empty
N/A
XJ14
N/A
Empty
N/A
TR1 - 5
8 mm open ended wrench
SEL
Brown
TR1 - 0
8 mm open ended wrench
0V
Brown
1