Inverter Board 300PL1 Neon Lights
2-4-3.
Inverter Board 300PL1 Straps
2-4-4.
Inverter Board 300PL1 Test Points
LED
Function
Color
Behavior
L1
600 V Present
Clear
Normal status: ON
ON when the 600 V from the Mains Distribution
Board 200PL1 is present.
OFF when there is a problem with the 600 V from
the Mains Distribution Board 200PL1. In this case,
proceed as follows:
Check the states of the all the LEDs on the Mains
Distribution Board 200PL1(see
).
• If the states of any of the LEDs are abnormal, fix
• If the states of all the LEDs are normal, check with
a multimeter that the voltage between the X3 and
X4 connectors on the Inverter Board 300PL1 is at
415 V.
If the voltage is at 415 V, change the Inverter
Board 300PL1.
If the voltage is not at 415 V, check the cabling
between the Mains Distribution Board 200PL1
and the capacitors of the DC Bus on the Inverter
Board 300PL1.
Strap
Function
ST1
Gate TH1
ST2
Gate TH2
Test Point
Label (in French)
Function (in English)
PT1
A/TH1
PT2
HF1
PT3
K/TH2
PT4
600VF
PT5
0V/600VF
600 VF supply at 0 V