GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A014 - Connect Gantry to Control Station
Page no. 459
Chapter 6
JC-PHY-A-014.fm
Note:
The monitor DVI cable should already be connected via the DVI-I - display port adapter. However,
ensure that it is connected to the left hand side display port connector (labelled number 1) on the
back of the ADS computer.
4. All the other cables (i.e. monitor, keyboard, mouse/trackball, service key, Ethernet, and power)
should already be connected to the ADS Computer. Ensure that all these cables are still firmly in
place. Refer to Illustration 1 on
Note:
A trackball can be provided as an option in place of the mouse.
Do not
attempt to connect a mouse
and trackball simultaneously.
5. Connect the Ethernet cable labeled 2385966 from the harness (originating from the Gantry CPU) to
port 3 on the Ethernet switch.