Job Card ELE A010 - Set Lift Travel Limits
Page no. 567
Chapter 7
JC-ELE-A-010.fm
6-2
Calibration Procedure
6-2-1
Upper Optical Fork Positioning
•
Use a measuring tape to measure the height of the ceiling:
1. If the ceiling height is equal to or greater than 2500 mm
(98.4 inches) (measured from the floor),
-
Leave the optical fork in the default position (the
upper position, allowing 850 mm lift travel),
2. If the ceiling height is between 2400 mm (94.6 inches)
and 2500 mm (98.4 inches)
-
Remove the front covers (refer to
A044 - Remove/Reinstall Gantry Covers
-
Place the optical fork in the middle position (allows
750 mm lift travel).
3. If the ceiling height is between 2300 mm (90.6 inches)
and 2400 mm (94.6 inches)
-
Remove the front covers (refer to
A044 - Remove/Reinstall Gantry Covers
-
Place the optical fork in the lower position (allows
650 mm lift travel).
Note:
The ceiling height ranges quoted above assume that there is a 70 mm marginal space between
the top of the Tube Head and the ceiling.
6-2-2
Lower Limit Adjustment (Lift Potentiometer Lower Reference Voltage at 0 mm)
Use the following steps to set the lower lift travel limit (which corresponds to the Lift Potentiometer Lower
Reference Voltage):
1. Launch the Service Desktop (see
) then click the
Cali-
bration
button
.
2. From the
Calibration Home Page
that appears, click the
Positionner
button.
3. From the
Calibration: Positioner Summary
page
that appears, click the
Lift
button
.
4. From the
Lift Position and Upper Limit Calibration
page that appears, follow the on-screen instruc-
tions: set the arm rotation position to 0° and remove the Bucky.Then press the
Start
button to start
the calibration process.
The Gantry LCD updates with the following message:
Adjust lower limit. Lower arm until you reach min usable height defined in JC ELE A010. Then press
Ceiling Height:
____ mm
100
100
Fork
position
set
by
default
at
installation
750
mm
lift
travel
850
mm
lift
travel
650mm
lift
travel
See