3-1-4
Nephtys M410 (IS4-3 FMI 12075)
•
The AWS Application Software for the V3 Control Stations (U20 Workstation) was updated from
V53.0 to V53.10.1.
•
The AWS Application Software for the V1/V2 Control Stations (SB150 Workstation) was updated
from V32.10 to V43.10.1.
•
The Gantry CPU board was updated from NSC.1.0.10 to NSC.1.1.7.
•
The Interface board was updated from FPGA 2E to FPGA 4A.
•
FMI 12075 (DS) installed on the install base.
3-1-5
Nephtys M411 (IS4-3 Fwd Prod)
•
Introduction of Sisfle Rotative Arm (replaces Ergotron B Rotative Arm) on the Control Station.
•
The Gantry CPU board for manufacturing was updated from NSC.1.1.7 (5233827-2) to NSC.1.1.8
(5233827-3).
•
Introduction of V2 Clutch, which is only compatible with the Rotation Board (2375488-12). However,
the Rotation Board (2375488-12) is compatible with both V1 and V2 Clutches.
Note:
The Rotation Board (2375488-11) is not a FRU. If you need to replace the Rotation Board
(2375488-11), order Rotation Board (2375488-12).
•
A new Gantry Update CD-ROM (5174994-7) is delivered with the system.
Note:
The Gantry Update CD-ROM (5174994-7) is not a FRU. If you need to replace the Gantry CD-
ROM, order Gantry Update CD-ROM (5174994-6).
3-1-6
Nephtys M412 (Orion 1)
The support of 5 new languages and the replacement of Brazilian Portugese by continental Portugese,
which results in the introduction of:
•
a new Generator CPU Board (2385602-9), which supports additional languages.
•
a new Gantry CPU Board (5233827-4), which supports additional languages.
•
ADS 53.10.10 on V3 Control Stations, which supports 11 languages in the user interface, and 19 lan-
guages in the embedded user documentation.
•
a new Lift Board (2373567-9), which supports output of additional characters to the Gantry LCD
Readout.
Other product enhancements, including the introduction of:
•
a new Paddle Holder Board (2369391-2), which has DIP switches and better reliability. The Paddle
Holder Board (2369391-2) is fully backward comptible.
•
new Detector FRUs (2383172-4 and 2383172-5).
•
Type 2 Lift Screw Assembly. The Type 2 Lift Screw Assembly differs from the Type 1 Lift Screw
Assembly, such that it uses a different Ball Bearing Ring, and its associated Lift Motor has an earth
cable.
3-1-7
Nephtys M412 Step1
•
Introduction of X-ray Console (2282675-7) and new Arm Keypads (5268425).
3-1-8
Nephtys M413
•
Introduction of one Gantry board FRU strategy.
•
Safety compliance : New tools and filmless procedure available for Collimator checks.
•
Introduction of a stronger 3000 N Rotation Gas Spring (5154719).
•