GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A042 - Bar Code Scanner (Option)
Page no. 723
Chapter 7
JC-ELE-A-042.fm
6-7
Complete a Medical Procedure Card
Perform the following procedure to check for correct installation and configuration of the scanner.
1. Power up the AWS and when prompted at the command line
Login:
prompt, login as the sdc user as
follows:
a. At the
Login:
prompt, enter:
sdc
b. At the
Password:
prompt, press the
<Return>
key.
2. At the AWS login screen, select
clinical
from the
Username
drop-down field, type
clinical
in the
Password
field and click the
OK
button.
3. Wait until the Browser to appear.
4. Request a new exposure and open the Medical Procedure Card (MPC).
5. Click the first text input field
Last Name
with the left mouse button to set the keyboard focus.
6. Complete the MPC by scanning all the symbols given on the following page in order, scanning left-to-
right/top-down. As the MPC user interface can change from one AWS release to another, the layout
and entries given in these symbols are templates only: field values are not meaningful and can have
any value. To go to (change focus to) the next input field, scan the $I bar code symbol.