Chapter 9
Page no. 860
ERR-SUB-A-003.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
102/030
DEBUG
ONLY
LIFT POSITION
ERROR
There is a position error. The motion did not stop at
the target/limit position and went further than the
allowed margin.
Refer to
JC DIAG A1014 LIFT POSITION OVER-
to fix the problem.
A1014
102/031
WARNING
LIFT DECELERA-
TION ERROR
There is a deceleration error. The motion began to
slow down but did not stop before the allowed time.
Refer to
JC DIAG A1014 LIFT POSITION OVER-
to fix the problem.
A1014
102/032
HALT
LIFT MOTION
STARTED IN
REVERSE DIREC-
TION
The motion was about to go in the reverse direction
to what was requested.
Change the Lift board then perform Lift Calibration.
D/R A201
CAL A031
102/033
HALT
LIFT DSP TIMEOUT The Lift DSP did not answer to the MCU before the
timeout.
Change the Lift board then perform Lift Calibration.
D/R A201
CAL A031
102/034
WARNING
LIFT CAN COM-
MAND NOT PRO-
CESSED
CAN command not processed. The Gantry CPU sent
command to the lift node before the lift node could
acknowledge the previous command.
Reboot the system from the X-ray Console.
Retry the command.
If the problem persists, contact your OLC.
102/035
HALT
LIFT RX CAN BUF-
FER OVERRUN
There has been a CAN overrun. The input queue of
the incoming CAN messages of the MCU is full and
the message dropped.
Should not occur in field. If this error occurs contact
your OLC.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card