ERR SUB A003 - Gantry Error Messages
152/021
DEBUG
ONLY
PDU MCU reseted
while System is ON
The PDU MCU was reset while the system was on.
Ignore this error. If the error persists, contact your
OLC.
152/022
DEBUG
ONLY
A/D converter mod-
ule error on PDU
board
The A/D converter on the PDU board (U13) is bad
and the measurements are invalid.
Ignore this error. If the error persists, contact your
OLC.
152/023
DEBUG
ONLY
Diagnostic channel
selection circuit fail-
ure
PDU internal error
Ignore this error. If the error persists, contact your
OLC.
152/024
DEBUG
ONLY
Internal software
error
There has been an internal error with the PDU soft-
ware.
Ignore this error. If the error persists, contact your
OLC.
152/025
DEBUG
ONLY
Temperature is too
high or too low
The MPCB box temperature is out of the specified
+9 to +60 degrees operating range.
Ignore this error. If the error persists, contact your
OLC.
152/026
HALT
Low power 2 fuse
failure
The low power 2 fuse (F1) to the nodes is faulty.
See
JC DIAG A1504 LOW POWER 2 FUSE IS BAD
to fix the problem.
A1504
152/027
HALT
High power fuse fail-
ure
The high power fuse (F2) to the motors is faulty.
See
JC DIAG A1505 HIGH POWER FUSE IS BAD
to fix the problem.
A1505
152/028
HALT
Low power 1 fuse
failure
The low power 1 fuse (F3) to the nodes is faulty.
See
JC DIAG A1506 LOW POWER 1 FUSE IS BAD
to fix the problem.
A1506
152/029
DEBUG
ONLY
CPU fuse failure
The CPU fuse (F4) is faulty.
See
JC DIAG A1507 MAIN CPU FUSE IS BAD
to fix the problem.
A1507
152/030
HALT
Generator on off
relay failure
Generator on/off switch relay (or its driver) is faulty.
The relay that turns off the Generator at shutdown
has over/under current on its drive coil.
Check the following relays:
Relay-X1 16 A
Relay-K1 K2 8 A
Relay-18A 230 V