ERR SUB A003 - Gantry Error Messages
Page no. 913
Chapter 9
ERR-SUB-A-003.fm
161/023 ADVISORY
LCD: A06 Lift end
position is reached
The lift UIF request was rejected by the Gantry CPU
because it is already at its end position.
Operator error. No service recommended actions.
161/024 ADVISORY
LCD: A05 Compres-
sion upper position is
reached
The compression UIF request was rejected by the
Gantry CPU because it is already in its topmost posi-
tion.
Operator error. No service recommended actions.
161/025 ADVISORY
LCD: A09 Max
power driven com-
pression force is
reached
The maximum power-driven compression force has
been reached.
Operator error. No service recommended actions.
161/026 ADVISORY LCD: A08 Max man-
ual compression
force is reached
The maximum manual compression force has been
reached.
Operator error. No service recommended actions.
161/027 ADVISORY
LCD: A14 Risk of
collision with floor
An active lift motion was stopped or a lift UIF request
was rejected because a potential collision situation
was detected by the Gantry CPU.
Operator error. No service recommended actions.
161/028 ADVISORY
LCD: A14 Risk of
collision with floor
A rotation lift motion was stopped or a rotation UIF
request was rejected because a potential collision
situation was detected by the Gantry CPU.
Operator error. No service recommended actions.
161/030
HALT
COMPRESSION
CONTROL FAILURE
The compression node did not start. At power-up,
the compression node sends an acknowledgment
(which can be different from the first heartbeat sig-
nal) to Gantry CPU. If the compression node does
not send an acknowledgment, the Gantry CPU cre-
ates this error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
161/031
HALT
LIFT CONTROL
FAILURE
The lift node did not start. At power-up, the lift node
sends an acknowledgment (which can be different
from the first heartbeat signal) to Gantry CPU. If the
lift node does not send an acknowledgment, the
Gantry CPU creates this error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action