Error messages
I50 Lock Jammed
The system is in degraded mode:
Gantry device configuration is not
a standard configuration. The
system functionality is not opti-
mum.
1) If the installed paddle is not a Senograph DS
paddle:
A. Remove it.
B. Install a Senograph DS paddle.
2) If the installed paddle is a Senograph DS pad-
dle:
A. Remove it and reinstall it.
B. If the warning persists, check with another
Senograph DS paddle.
B1. If the warning disappears with another
Senograph DS paddle, the paddle that generated
the warning is not recognized by the system; the
thickness/force measurements with this paddle are
not accurate.
B2. If the warning persists with another Seno-
graph DS paddle, the thickness/force measure-
ment with all paddles are not accurate.
I51 Lock Overheated
The system is in degraded mode.
The tube head fan has failed. As
a result, tube cooling is not effec-
tive and the tube thermal algo-
rithm could not prevent exposure
abort due to tube overheating.
1) The system can continue to be used but please
reduce the cadence of use.
Add time between exposures and exams to
allow the tube to cool.
2) Contact GEMS Field Service to repair the tube
head fan.
I90 Auto-Decompres-
sion is Off
The automatic
decompression mode has been
inhibited by the user.
If auto-decompress mode is wanted, switch modes
by changing the parameter in the Setup/Medical
menu. Otherwise no action is required; this infor-
mation is normally displayed.
S01 Aborted exposure
Exposure in progress has been
aborted because the Exposure
Button was released by operator
before the end of exposure.
Power OFF/ON from the X-ray Console.
Start exam again.
S02 Exposure aborted
by HV arcing
Exposure was aborted because
more than eight X-ray tube arcing
events occurred.
Power OFF/ON from the X-ray Console.
Start exam again.
S05 Wrong selection of
kV/track couple
Selected kV/focal track is not
enabled (kV values are too low in
the Rh focal track).
Change the configuration.
S07 Exposure aborted
by IDC
The exposure was aborted by the
Image Detection Controller.
1. Power OFF/ON from the X-ray Console.
2. If the problem occurs again at the next acquisi-
tion, power OFF the whole system and wait for the
end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
S08 Required power is
too high
Power requested in Manual
Mode is too high.
Reduce kV or mAs values.
Message on
X-ray Console
Explanation
Corrective Action