Chapter 3
Page no. 102
S2100D How To.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Senographe Operations - How to...
ILLUSTRATION 2 -
APPLICATION MODE
CHECKSUM OF SAVED PARAM
CKSUM
CLEAR
GENERATOR CONFIGURATION
PRES.
MASMAX
TUBE
PMAX
T_MAX
RINIT
MAX EXPOSURE TIME T(s)=(value)
MAX AUTHORIZED POWER POWE = (value)
MAX MAS ALLOWED mAs=(value)
TYPE OF TUBE 2foc/2met/1ang
3.5KW
4.0KW
SMSA
4.5KW
FOCUS
3.0KW
VALID
ANODE
NEXT
5.0KW
DUAL
PRS-B
SUB-SYSTEMS PRESENCE
SMDA
DM
SINGLE
NEXT
VALID
TYPE OF ANODE 2Metal
FOCUS SELECTION bifocus
CLEAR ALL SAVED PARAM
VALID
WARNING
USING THE CLEAR FUNCTION WILL
ENTAIL A COMPLETE RE-CALIBRATION
OF THE GANTRY, GENERATOR AND AEC.
WARNING
RE-INITIALIZING THE TUBE
WILL ENTAIL RE-CALIBRATING
THE TUBE HEATER CURRENT.
USE THE FUNCTION ONLY
WHEN YOU HAVE REPLACED
THE X-RAY TUBE.
PRS-DS
PRS-IDC
YES
ARM PRESENCE
NO
DIGITAL STATION PRESENCE
IMAGE DETECTOR CTRL PRES.
YES
NO
YES
NO
MODEL OF TUBE : (Maxray 100TH-M1)
X-RAY DISCHARGE Delay (ms)
GS412
70TH-F
NEXT
VALID
70TH-D
MODEL
DCHARG
GS412
VALID
70TH-F
VALID
70TH-D
VALID
100TH-M1
100TH-M1
VALID