GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A189 - Interface Board
Page no. 1125
Chapter 9
JC-DR-A-189.fm
Job Card D/R A189 - Interface Board
Chapter 9
1
SUPPLIES
2
TOOLS
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 40 minutes
4
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
•
When handling circuit boards, take precautions against damage by ESD (electro-static discharge).
5
PREREQUISITES
None
6
PROCEDURE
6-1
Removing the Interface Board
1. Remove the Gantry CPU board from the Interface board according to
Job Card D/R A190 - Gantry CPU Board
.
2. Disconnect all the cables and optical fibers from the Interface board and make a mental note where
they connect to on the Interface board. Ensure that you do not move them too much so it is easier to
re-connect them later on.
CAUTION
Disconnect with caution to avoid damage to the connectors and cables.