Service Information and Procedures Class A 2385072-16-8EN
Senographe Operations - How to...
Page no. 115
Chapter 3
S2100D How To.fm
GENE-INSTALL
DOSE
AOP
LF
DOSE : Select Focus
SF
MoMo
DOSE : Select Spectrum LF
MoRh
RhRh
CALIB
DOSE : LF Mo/Mo CALIB
PARAM
0
DOSE : LF Mo/Mo CALIB
0.2
0.5
CALCUL
DOSE : LF Mo/Mo 0 mm Al
Meas.D
CALIB
NEXT
DOSE : LF Mo/Mo 0.0 mm Al (mGy) value
VALID
DOSE : LF Mo/Mo : HVL ( 2 values)
Same menu
Idem
Idem
Idem
THIS MENU CONSISTS OF PRO-
DUCTION TEST PROCEDURES
THAT ARE OF NO USE TO THE
SERVICE ENGINEER.
LF
Structure as
CHANGE
NEXT
DOSE : LF Mo/Mo (Name) (Value)
VALID
WARNING
THE VALUES OF HVL AND TUBE
YIELD ARE COMPUTED DURING
THE CALIBRATION PROCEDURE
AND MUST
NEVER
BE MODIFIED
MANUALLY FOR ANY REASON
WHATSOEVER.
Same menu
MoMo
Structure as
MoAl
RhAl