GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A202 - Upper Optical Fork
Page no. 1167
Chapter 9
JC-DR-A-202.fm
Job Card D/R A202 - Upper Optical Fork
Chapter 9
1
SUPPLIES
2
TOOLS
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 25 minutes
4
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5
PREREQUISITES
•
Raise the arm to an intermediate position to facilitate the
access to both upper (1) and lower (2) optical forks.
•
Switch off the system power.
•
Remove the Front cover and the right column cover (Refer to
Job Card PHY A044 - Remove/Reinstall Gantry Covers
•
Loosen and remove the Gantry curtain (refer to
1
2