GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A208 - Gantry Curtains
Page no. 1193
Chapter 9
JC-DR-A-208.fm
Job Card D/R A208 - Gantry Curtains
Chapter 9
1
SUPPLIES
Kit containing upper and lower Gantry Curtains
2
TOOLS
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 60 minutes
4
SAFETY PRECAUTIONS
WARNING
While performing this procedure, you must:
Power off the Gantry.
Wear safety glasses and gloves.
5
PREREQUISITES
•
Switch off the Gantry power.
•
Remove the front and column covers (refer to
Remove/Reinstall Gantry Covers
.
•
Switch on the Gantry power.
•
Use the lift command buttons to move the arm vertically until the two
springs (1) are fully visible in the opening at the rear of the column
frame.
•
Switch off the Gantry power.
1