7.
HOW TO USE DIAGNOSTICS
Click the
Diagnostics
button
in the Service Desktop navigation bar to launch diagnostic tests for the
different subsystems and communication links of the Senographe system.
Specific instructions are given for each diagnostic test, and results are displayed at the end of each test.
! Notice:
After running diagnostics, always reboot the system before handing it back to the customer, to en-
sure correct configuration. This is especially important after tests which require disconnection of
communications links.
ILLUSTRATION 14 - DIAGNOSTICS ON V1/V2 CONTROL STATIONS