GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A243 - Compression Potentiometer and Cable
Page no. 1289
Chapter 9
JC-DR-A-243.fm
Job Card D/R A243 - Compression Potentiometer and Cable
Chapter 9
1
SUPPLIES
Compression Potentiometer and Cable (with Support)
2
TOOLS
•
•
Mammography compression scale with digital display
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 1 hour (incl. calibrations and functional checks)
4
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5
PREREQUISITES
•
Rotate the arm to 0°.
•
Lower the arm until you have a good access to the Compression Potentiometer.
•
Switch off the Gantry power.
•