GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A248 - Force Sensor
Page no. 1313
Chapter 9
JC-DR-A-248.fm
Job Card D/R A248 - Force Sensor
Chapter 9
1
SUPPLIES
2
TOOLS
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 60 minutes
4
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5
PREREQUISITES
•
Switch off the Gantry power.
•
Remove the Paddle.
•
Remove the Paddle Holder cover (refer to
PHY A044 - Remove/Reinstall Gantry Covers
) for access to the Force Sensor (1)
1