Chapter 3
Page no. 134
S2100D How To.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Senographe Operations - How to...
9-6. Calibration of Rotation
The arm rotation calibration sets travel limits in software, defining the maximum positive and negative
usable angles.
Refer to
Job Card CAL A021 - Arm Rotation Calibration
9-7. Calibration of Compression Force
The compression force sensor is factory calibrated. However, if parts have been replaced or disturbed
which affect the software or hardware settings, calibration must be carried out.
Refer to
Job Card CAL A048 - Calibration of Compression Force Sensor
.
9-8. Compression Thickness Calibration
The breast thickness calibration calibrates the absolute breast thickness measurement read by a linear
potentiometer linked to the compression carriage. This is done by defining two points in the range of
travel of the compression carriage.
Refer to
Job Card CAL A050 - Compression Thickness Calibration
.