GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A288 - Collimator Lamp
Page no. 1383
Chapter 9
JC-DR-A-288.fm
Job Card D/R A288 - Collimator Lamp
Chapter 9
1
SUPPLIES
New centering device lamp (only genuine part number 2377836 must be
used, to ensure correct light intensity, easy filament centering and reliability).
2
TOOLS
including a small Pozi screwdriver
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 2 hours
4
SAFETY PRECAUTIONS
Not applicable.
5
PREREQUISITES
•
Rotate the arm to 0° and move it to its comfortable working
position to access to the Collimator Lamp (1).
•
Switch off the Gantry power.
•
Remove the tube cover (refer to
Remove/Reinstall Gantry Covers
) to access
the Collimator (1).
1