GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A049 - Compression Paddle Offset Calibration
Page no. 1711
Chapter 9
JC-CAL-A-049.fm
Job Card CAL A049 - Compression Paddle Offset Calibration
Chapter 9
1
SUPPLIES
None
2
TOOLS
Set of acrylic plates
Calipers
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 5 minutes per Compression Paddle (approximately 40 minutes for a set of Compression Paddles)
4
SAFETY PRECAUTIONS
No specific safety precautions are applicable.
5
APPLICABILITY
This Job Card should be used if you witness problems when peforming
and you have a Gantry CPU Board running NSC 1.2.13 or above.
6
PREREQUISITES
The procedure to calibrate the final offset for each Compression Paddle uses a command, whose syntax
refers to the
padlle number
. Before proceeding, consult Table 1 to know which paddle number corre-
sponds to each of the Compression Paddles delivered with the system.
TABLE 1 - PADDLE NUMBER REFERENCES
Paddle
Number
Products
Compression Paddle Name
(Ref. Number on Label)
Photo
1
DS
Flexible 19x23
(2383182-2)
2
DS
Standard 19x23
(2373344)