Chapter 3
Page no. 270
Indicators and Switches.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Central Listing
2-7. Supply Command Board 200PL2
2-7-1.
Supply Command Board 200PL2 Connectors
Connector
Function
XJ1
To the XJ1 connector on the Generator Command Board 400PL1 (standby power supply)
XJ2
To/from the XJ11 connector on the Mains Distribution Board 200PL1 (12 V, control/data signals)
XJ3
Fan (+27 V)
XJ4
To the XJ4 connector on the Relay Board 200PL4 (control/data signals)
XJ5
To the XJ2 connector on the Generator Command Board 400PL1 (0, +250 V DC)
To the CN1 connector on the BT Power Supply (0, +250 V DC)
To the XJ3 connector on the Anode Starter Board 300PL2 (0, +250 V DC)
XJ6
+ 27 V on the C1 capacitor in the Generator
XJ7
0 V on the C1 capacitor in the Generator
XJ8
Thermostat (+27 V)
XJ9
To auto transformer (250 V)
XJ10
To emergency stop button on Gantry and thermal switch the Auto Transformer (security signal)
XJ11
To the XJ13 connector on the Generator Command Board 400PL1 (control signals)