Chapter 3
Page no. 300
Cables.FM
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Cable Lay-out and Pin-out
1-7. W104 - PDU Power Cable
PDU Board PL101 J1 to EMC filter E101 P’, N’.
1-8. W105 - 48 V DC Cable
AC/DC Converter Module J2 to EMC Filter E102 N.
1-9. W106 - AC/DC Power
AC/DC Module Board PS101 J137 to EMC filter E101 P’, N’.
1-10. W107 - AC/DC Databus
AC/DC Module Board PS101 J138 to PDU Board PL101 J3.
J2-12
L Power On
Command/reserved
J2-13
H-Motion Enable
RT-CAN Diff 5/2.5V
J2-14
L-Motion Enable
RT-CAN Diff 2.5/0V
J2-15
Command CAN Electric
CAN Ground
J2-16
P Power supply
12 V
J2-17
H-CAN Network
CAN Diff 5/2.5V
J2-18
L-CAN Network
CAN Diff 2.5/0V
J2-19
H-Spare
RT-CAN Diff 5/2.5V
J2-20
L-Spare
RT-CAN Diff 2.5/0V
J1
E101
1
4
P’
N’
Shielding
J137
E101
Pin 1 : Phase
Pin 3 : N
1
3
W 106
2345983