9.
PLANNING FOR STORAGE
9-1. Temporary Storage in the Hospital
! Notice:
There is normally a short delay (e.g., overnight) between the delivery of the equipment and its in-
stallation.
If this delay is not short (more than two days), it is essential that a suitable storage room is
available to receive the equipment in its crates
.
for information on the environment required.
9-2. Packing Information
The table below lists the main dimensions and masses of shipping crates.
TABLE 6 - SHIPPING DIMENSIONS AND MASSES.
Item
Dimensions in mm
(inches)
Mass in
kg
(lbs)
Depth
Width
Height
Crate1
2066
(
81.34)
848
(
33.39
)
2292
(
90.24
)
641
(1410.2)
Crate 1 includes the Gantry and the Generator Cabinet and the ramp for removing the Gantry, Generator and
Control Station
Crate 2
1110
(
43.70
)
770
(
30.31
)
1850
(72.83)
210.5
(464.07) for
truck transportation
without side covers
and lid
223.5
(492.73) for
plane/boat trans-
portation with side
covers and lid
Crate 2 includes the Control Station
Crate 3
872
(
34.33
)
2270
(
89.37
)
260
(
10.24
)
104.4
(
229.68
)
Crate 3 includes the box containing the Detector. (20 kg, D=766 (30.16) , W= 840 (33.07), H =654 (25.39) and
the Radiation Screen in its protective packaging
Crate 4
720
(
28.35)
1300
(
51.18
)
1975
(
77.76
)
147
(
323.4
) *
Crate 4 includes the Accessories Cabinet, which includes the LCD Monitor, X-ray Console and all the Accesso-
ries (boxed) . * - The contents can vary, and the mass can be slightly different to 147 kg.