GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Pre-Installation System Requirements
Page no. 341
Chapter 4
Chap-Pre-Requirements.fm
9-3. Constraints for Moving the Equipment Into the Room
The minimum dimension of the entry door to move in the (uncrated) Gantry on its wheels with a 61.5 mm
(2.40 inches) radius are:
•
door opening at least 700 mm (27.6 inches) wide.
•
height of 2020 mm (79.53 inches) with Gantry covers.
•
height of 1897 mm (74.69 inches) without Gantry covers.
Note:
If the hospital doors are less than height of 2020 mm (79.53 inches) high, prepare time to remove
the Gantry covers during the delivery of the Senographe System. You will need to remove the Gan-
try covers so that you can move the Gantry under the doors.
Note:
In cases where a Senographe system is not installed on the ground floor of a building, you must
consider the size of the hospital elevators. In cases where hospital elevators are small, you will
need to disconnect the Generator from the Gantry and move them separately. The minimum depth
of hospital elevators must be slightly larger than 1273 mm (50.1 inches) in order to be able to move
the Gantry.
10. ROOM LAYOUT PLANNING
10-1. Dimensions and Masses
Refer to the following pages for more information on major components.
TABLE 7 - SYSTEM COMPONENT DIMENSIONS AND MASSES
.
Component
Depth
mm
(inches)
Width
mm
(inches)
Height
mm
(inches)
Mass
kg
(lbs)
Gantry
(with Stereotaxy Positioner)
1273
(50.1)
616
(245.25)
Required space:
1842
(72.5)
1930
(76)
to
2430
(95.67)
418
(921.5)
Generator Cabinet
437 (
17.2)
640 (
25.2)
1330 (
52.5)
180 (
396.8)
19" monitor (default)
98
(
3.9
)
410
(
16.15
)
340
(
13.4
)
8
(
17.6
)
3 MP 21.2" monitor (option)
97
(3.8)
500
(19.7)
376
(14.8)
7.4 (16.3)
X-ray Console
180 (
7)
550 (
22)
85 (
3.5)
1.6 (
3.5)
Radiation Shield (Lemerpax)
9
(
0.35
)
700
(
27.5
)
1479
(
58.5
)
26
(
57.3
)
Radiation Shield (Mavig)
6
(
0.24
)
700
(
27.5
)
1150 (45.3)
15 (33.1)
Control Station
without
radiation
shield, monitor, and X-ray Con-
sole
400 (
15.75)
710 (
27.9)
2225 (
87.5)
181.2
(399.5)