Chapter 4
Page no. 342
Chap-Pre-Requirements.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Pre-Installation System Requirements
ILLUSTRATION 2 - GANTRY DIMENSIONS - MM (INCH)
•
Anchor Points for the Gantry
Refer to
1930
(76)
All measurements shown as:
mm
(inches)
2447
(96.4)
1273
(50.1)
1842
(72.5)
Gantry base
1842
(72.5)
630
(24.8)
788
(31.0)
X
Y
Z
Center of Gravity (CG):
X = 325 (12.8)
Y = 3 (0.12)
Z = 1024 (40.3
)
Z
CG
CG
plate
2430 (94.67)
Maximum
Height