GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A030 - Generator and Gantry Disconnection Reconnection
Page no. 483
Chapter 6
JC-PHY-A-030.fm
7
GENERATOR DISCONNECTION PROCEDURE
Use the following steps to disconnect the Gantry connections from the various parts of the Generator:
1. Remove the four Generator side panels (see
Job Card PHY A042 - Remove/Reinstall Generator
) to reveal the inside of the Generator.
2. Remove the cable ties (1) then from the Generator Interface Board 400 PL2, disconnect the cables
from the XJ4 and XJ10 connectors.
XJ4
XJ10
1