GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A044 - Remove/Reinstall Gantry Covers
Page no. 523
Chapter 6
JC-PHY-A-044.fm
Job Card PHY A044 - Remove/Reinstall Gantry Covers
Chapter 6
1
SUPPLIES
None
2
TOOLS
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 20 minutes to remove covers, and 30 minutes to reinstall covers
4
SAFETY PRECAUTIONS
WARNING
Before beginning work, shut down the equipment and switch off the Generator hospital
supply circuit breaker. Lock it in the off position or mark it clearly with a warning sign to
prevent others from applying power.
Never lay a cover down with the painted side on a hard surface.
5
PREREQUISITES
None
6
OVERVIEW
For an overview of the covers, refer to
List of covers:
•
-
Cable Exit Cover
-
Top Column Cover (Left and right)
-
Bottom Column Covers (Rear and front)
-
Front covers (Left and right)
•
Compression and Arm Covers (section
-
Compression Carriage Cover
-
Compression Arm Covers (Left and right)
•
Tube Head Covers (Left and right) (section
)
•
Detector Covers (Upper and lower covers) (section
There are different types of covers
•
Pilot covers
•
Series covers which are divided into two kinds:
-
Flexible covers
-
Rigid covers
The procedure is valid for all covers except where specified.