GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A043 - AWS Monitor Calibration
Page no. 751
Chapter 7
JC-ELE-A-043.fm
8-6-2
19" Monitor and V1/V2 Control Station
1. The SMPTE pattern is already displayed on the AWS Monitor. If not, refer to
to display
the SMPTE pattern.
2. If necessary, unlock the OSD (see
) and move the OSD to a convenient position (see
3. Press the
Key1
button to access the OSD main menu, then keep pressing the
Key1
button until
Auto
functions
is selected, and press
the
Key2
button
to enter the
Auto functions
menu.
4. In the
Auto functions
menu, keep pressing the
Key1
button until
Execute selected auto functions
is
selected, and press
the
Key2
button
to perform an auto calibration.
The auto-calibration takes place.
5. If necessary, unlock the OSD (see
) and move the OSD to a convenient position (see
6. Press the
Key1
button to access the OSD main menu, then keep pressing the
Key1
button until
Brightness / Contrast
is selected, and press
the
Key2
button
to enter the
Brightness / Contrast
menu.