Scenario PDR A001 - Post-D/R Procedures
2
POST-D/R PROCEDURES
2-1
Control Station
•
If
Stereo or
2D Biopsy Optical Localizer option exists, calibrate the LCD Monitor Monitor
2-1-2
•
After installing or restoring IDC software, IDC parameters must be restored as described in
ELE A052 - Back-up/Restore Procedure
. If the restore is successful, it is possible to use
the system normally without further intervention.
•
If the restore is not possible or is not successful, the following calibrations and checks must be per-
formed:
Calibrations - Checks
Job Card
Calibrate backlight
Job Card ELE A043 - AWS Monitor Calibration
Calibrations - Checks
Job Card
Bad Pixel Calibration
Job Card CAL A041 - Bad Pixel Calibration
Gain calibration
Job Card CAL A043 - Detector Gain Calibration
flat field tests
Job Card ELE A032 – IQ Tools Flat Field Test
If IQ tests not OK, repeat all calibrations from Bad Pixel
Half Value Layer
Job Card CAL A046 - Half-Value Layer Measurement
Conversion Factor
Job Card CAL A045 - Conversion Factor Measurement
Use the same dosimeter as that used for CAL A046
AOP Calibration
Job Card CAL A051 - AOP Calibration
mAs non linearity
Job Card CAL A053 - mAs Non-linearity Calibration
AOP/SNR Check
Job Card ELE A038 - AOP Mode and SNR Check
If AOP/SNR check not OK, repeat all calibrations from AOP Calibration