GE Analytical Instruments ©2010
84 of 226
DLM 74001-04 Rev. A
Chapter 4: Basic Analyzer Operation
*All Coils (write) and Input registers are 16-bit unsigned.
Notes:
Values are IEEE 32-bit floats split into two 16-bit registers.
Firmware Version (Engineering)
4
3017
*
Current Mode
4
3099
*
Reading Time - Year
4
3100
*
Reading Time - Month
4
3101
*
Reading Time - Day
4
3102
*
Reading Time - Hour
4
3103
*
Reading Time - Minute
4
3104
*
Reading Time - Second
4
3105
*
Error Type
4
3200
*
Error Time - Year
4
3201
*
Error Time - Month
4
3202
*
Error Time - Day
4
3203
*
Error Time - Hour
4
3204
*
Error Time - Minute
4
3205
*
Error Time - Second
4
3206
*
TOC Value Low Word
4
3300
*
TOC Value High Word
4
3301
*
TOC Units
4
3302
*
TC Value Low Word
4
3310
*
TC Value High Word
4
3311
*
TC Units
4
3312
*
IC Value Low Word
4
3320
*
IC Value High Word
4
3321
*
IC Units
4
3322
*
Sample Temp. Cond. Value Low Word
4
3500
*
Sample Temp. Cond. Value High Word
4
3501
*
Sample Temp. Conductivity Units
4
3502
*
Raw Sample Cond. Value Low Word
4
3510
*
Raw Sample Cond. Value High Word
4
3511
*
Raw Sample Conductivity Units
4
3512
*
Sample Temp. Value Low Word
4
3700
*
Sample Temp. Value High Word
4
3701
*
Sample Temperature Units
4
3702
*
Holding Registers
(read/write)
not used
Table 5: Modbus Register Table
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...