background image

29-0047-78 AC 04-2014

2

1.  Health and Safety 

3

  1.1. Scope 

 

1.2.  Intended use 

 

1.3.  Safety Notices 

 

1.4.  General Safety 

 

1.5.  General Hazards 

 

1.6.  Unpacking and Installation 

5

2.  Operating and Maintenance 

 

2.1.  Preparation before starting 

 

2.2.  Performing a run - cuvettes 

 

2.3.  Performing a run – Micro Volume Instruments 

 

2.4.  Post run procedures 

 

2.5.  User Maintenance 

 

2.6.  Novaspec III/Plus Lamp Change 

10 

 

2.7.  Ultrospec 8000/9000 Lamp Change 

10

3. Troubleshooting 

11 

 

3.1.  Emergency Procedures 

12 

 

3.2.  Recycling Procedures 

12 

 

3.3.  Recycling of hazardous substances 

12 

 

3.4.  Disposal of electrical components 

12 

  3.5. Specifications 

13 

 

3.6.  Manufacturing information 

13

4. Legal 

14

Page finder

Summary of Contents for SimpliNano

Page 1: ...Spectrophotometer Health Safety Document including General Operating Instructions Original Instructions ...

Page 2: ... 2 Performing a run cuvettes 7 2 3 Performing a run Micro Volume Instruments 8 2 4 Post run procedures 9 2 5 User Maintenance 9 2 6 Novaspec III Plus Lamp Change 10 2 7 Ultrospec 8000 9000 Lamp Change 10 3 Troubleshooting 11 3 1 Emergency Procedures 12 3 2 Recycling Procedures 12 3 3 Recycling of hazardous substances 12 3 4 Disposal of electrical components 12 3 5 Specifications 13 3 6 Manufacturi...

Page 3: ...onto the measurement area Typical applications of spectrophotometers include DNA RNA Oligo concentration and purity measurements as well as protein concentrations This manual addresses the needs of scientists and technicians who operate UV Visible spectrophotometers The level of information presented in this manual assumes the user possesses basic laboratory and technical skills and has the knowle...

Page 4: ...clude an appropriate decontamination certificate 1 4 General Safety This equipment has been designed to conform to the following directives 2006 95 EC Low voltage directive LVD 2004 108 EC Electromagnetic Compatibility EMC directive 2012 19 EU Waste Electrical and Electronic Equipment directive recast WEEE Recast 2011 65 EU Restriction on the use of certain hazardous substances ROHS directive 2006...

Page 5: ...y be connected to a separate PC To preserve the integrity of the measuring equipment it is essential that the attached PC itself conforms to basic safety and EMC standards and is set up in accordance with the manufacturers instructions If in doubt consult the information that came with your PC In common with all computer operation the following safety precautions are advised To reduce the chance o...

Page 6: ...ng capacity SimpliNano units 100 240 V AC 50 60 Hz 3 A UK plug only NanoVue Plus units 90 250v 50 60Hz Max 30 VA 3 A UK plug only The instrument should be positioned so that the power cable may be readily removed in the event of a hazard or malfunction occurring Site the instrument in an atmosphere free from dust and corrosive fumes Ultrospec 2100 instruments have the on off switch on the back of ...

Page 7: ...ntration purity The options within the application set up the instrument parameters such as wavelength and define any calculations taking place on the measured results 2 2 Performing a run cuvettes On the Ultrospec 10 Novaspec III Plus GeneQuant 100 1300 and Ultrospec 2100 7000 8000 9000 units samples are measured in a cuvette A suitable cuvette must be used for the application to be carried out i...

Page 8: ... and the Run Icon pressed If a single cell is being used e g a microcell it should be filled with reference solution placed in the front cell holder the Zero icon pressed the cuvette filled with sample placed in the front cell holder and the Run Icon pressed Typical results screen DNA application GeneQuant 2 3 Performing a run micro volume instruments On the SimpliNano units samples are measured i...

Page 9: ...ons or dilute detergent solutions but should always be rinsed in deionised water SimpliNano wipe the sample port with a dry lint free tissue between samples and at the end of the run this is sufficient to remove sample If more thorough cleaning is desired the port can also be cleaned with dilute 2 detergent solution followed by isopropyl alcohol NanoVue Plus wipe sample with tissue At end of sampl...

Page 10: ...ose of the faulty lamp in accordance with local regulatory practice 2 7 Ultrospec 8000 9000 Lamp Change 1 Switch the instrument off disconnect the power supply cord and allow lamps to cool 2 Locate the lamp access cover at the rear of unit and remove the two top screws and slide top cover back and lift off Never operate the unit with the lamp housing cover removed 3 4 Remove the lamp and dispose o...

Page 11: ...ot zero Cuvette placed in wrong orientation Wrong cuvette material used for wavelength Novaspec III Error code 001 symbol Lamp needs changing Novaspec III Error code 006 symbol Pixel clock too high Instrument needs optical alignment Contact service support Novaspec III Error code Flashing number Wavelength accuracy failure Instrument needs optical alignment Contact service support Novaspec plus X ...

Page 12: ...ual supplied on a CD with the instrument 3 1 Emergency Procedures In the event of a malfunction or hazard occurring the user responsible shall disconnect the unit from power and isolate the instrument for decontamination by an appropriate means if bio hazardous material is spilt on in or around the instrument Ultrospec 2100 instruments have the on off switch on the back of the instrument on the ri...

Page 13: ... 0 004 A at 1 A 0 006 A at 2 A at 440 nm 465 nm 546 1 nm 590 nm 635 nm 0 005 A using 60 mg l K2Cr2O7 within 0 5 of Absorbance value to 3 000 A at 546 nm 0 005 Abs or 1 of the reading whichever is the greater at 546 nm 2 0 or 0 010 A to 1 000A at 546nm whichever is the greater 0 05 A at 1 A using Neutral Density Filters Full specification and operating instructions are contained within the relevant...

Page 14: ...to the terms and conditions of sale of the company within GE Healthcare which supplies them A copy of these terms and conditions is available on request Contact your local GE Healthcare representative for the most current information http www gelifesciences com GE Healthcare UK Limited Amersham Place Little Chalfont Buckinghamshire HP7 9NA UK GE Healthcare Bio Sciences AB Björkgaton 30 751 84 Upps...

Page 15: ...29 0047 78 AC 04 2014 15 Page intentionally left blank ...

Page 16: ...l Avenue P O Box 1327 Piscataway NJ 08855 1327 USA GE Healthcare Japan Corporation Sanken Bldg 3 25 1 Hyakunincho Shinjuku ku Tokyo 169 0073 Japan For contact information for your local office please visit www gelifesciences com contact GE Healthcare Limited Amersham Place Little Chalfont Buckinghamshire HP7 9NA UK http www gelifesciences com imagination at work 29 0047 78 AC 04 2014 imagination a...

Reviews: