Ground Continuity Test
This test verifies that there is continuity (less than 100 m
Ω
resistance)
between all the exposed metal surfaces, which have the potential to
become energized, and the ground prong on the mains AC power cord.
Look for an exposed metal screw, or, if the metal surfaces are anodized or
painted, scrape off a small area in an inconspicuous area on the
aluminum casting, for the probe to make direct contact with the metal.
1. Connect the AC power cable to the GE T2100 Treadmill.
2. Use a digital multimeter to check all the metal surfaces of the
equipment as illustrated below. Make adjustments for any resistance
in the test leads.
If the measurements are significantly out of range, check for breaks in
the power cord or in the internal connections within the unit.
Connect Controlling Devices
1. Attach the appropriate controller cable from the stress system or
MTC-1. Use the supplied interface cable to connect the GE T2100
Treadmill to the host (i.e., CASE, MTC-1).
NOTE
The CASE and MAC 5000ST are connected to the RS 232 Serial
Port, and the MTC-1 is connected to the RS 422 Serial Port.
NOTE
On the CASE system, connect the GE T2100 Treadmill interface
cable in one of the following ports: COM 1, COM 2, COM C or
COM D. The CASE RSS modem requires the use of either COM 1
or COM 2. Move the GE T2100 Treadmill interface cable to COM
C or COM D only as necessary.
2. Verify that the emergency stop switch is connected.
3. Connect the AC power cord.
<0.1
Ω
020
Summary of Contents for T2100
Page 1: ...GE T2100 Treadmill Service Manual 2021403 031 Revision A ...
Page 8: ...vi GE T2100 Treadmill Revision A 2021403 031 ...
Page 9: ...Revision A GE T2100 Treadmill 1 1 2021403 031 1 Introduction ...
Page 10: ...1 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Page 19: ...Revision A GE T2100 Treadmill 2 1 2021403 031 2 Equipment Overview ...
Page 20: ...2 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Page 27: ...Revision A GE T2100 Treadmill 3 1 2021403 031 3 Installation ...
Page 28: ...3 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Page 45: ...Revision A GE T2100 Treadmill 4 1 2021403 031 4 Maintenance ...
Page 46: ...4 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Page 51: ...Revision A GE T2100 Treadmill 5 1 2021403 031 5 Theory of Operation ...
Page 52: ...5 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Page 73: ...Revision A GE T2100 Treadmill 6 1 2021403 031 6 Troubleshooting ...
Page 74: ...6 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Page 103: ...Revision A GE T2100 Treadmill 7 1 2021403 031 7 Assembly Disassembly ...
Page 104: ...7 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Page 133: ...Revision A GE T2100 Treadmill 8 1 2021403 031 8 Ordering Parts ...
Page 134: ...8 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Page 137: ...Revision A GE T2100 Treadmill A 1 2021403 031 A Appendix A Technical Specifications ...
Page 138: ...A 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...
Page 143: ...Revision A GE T2100 Treadmill B 1 2021403 031 B Appendix B Electromagnetic Compatibility EMC ...
Page 144: ...B 2 GE T2100 Treadmill Revision A 2021403 031 For your notes ...