4 Operation
4.3 Preparations before start
34
Typhoon
Operating Instructions
28-9623-56
AA
b) Place the screen on the Image eraser (A) and rest the screen on the lip (B) of the
Image eraser.
c)
To erase two small screens at once, suspend an unmounted screen from the
clip at the top of the Image eraser (C).
d) Turn the Image eraser power switch on (D).
e)
Set the Image eraser time by pressing the
Time
button (E).
•
Select
Normal
(F) position (out) for standard samples. The Normal setting
takes approximately 10 minutes.
•
Select
Extended
(G) position (in) if the background or residual image is high
(e.g., the original image contained readings of 104 counts and higher). The
Extended setting takes approximately 20 minutes.
f)
Press the
Erase
button (H).
3
Check for contamination of the screen between exposures or if the screen has not
been used for several days.
a) Clean and erase the screen as described above.
b) Store the screen in a clean, light-tight box.
Erase
Time
I m a g e E r a s e r
A
B
G
F
E
H
D
C
NOTICE
Do not use the clip on a TR screen. The clip can damage the surface of
the screen.