6 Troubleshooting
6.1 Power and communication
66
Typhoon
Operating Instructions
28-9623-56
AA
Typhoon 9400 and
9410 only:
The
Typhoon instrument is
on, but the external
Blue laser module is off
or the
Connection
light
on the instrument is
not lit.
The laser
communication
cable is not
connected between
the instrument and
the Blue laser
module, or the cable
is of wrong type.
Make sure to use the blue
Ethernet cable supplied with the
instrument.
The Blue laser
module is not
plugged in or the
wall outlet is faulty.
Check the connections and the
wall outlet.
Typhoon 9400 and
9410 only:
The
Laser
On
light on the Blue
laser module is not lit.
The laser is not
turned on.
Check the instrument state in
the
Typhoon Scanner Control
window. If it is in the
Sleep
state,
click
Initialize Scanner
to turn
on the laser.
Faulty laser.
If the light is not on 30 seconds
after you turn on the instrument
or click
Initialize Scanner
,
contact Technical Support.
Typhoon 9400 and
9410 only:
The Blue
laser module does not
turn off when you turn
off the instrument.
If the Blue laser module does not
turn off several minutes after
you turn off the instrument,
contact Technical Support.
Typhoon Trio and Trio+:
An error message is
displayed, stating:
Failed Communication
with Internal Laser
Unit.
Communication with
the internal blue
laser failed.
Turn off the power to the
instrument, wait a few seconds,
and the turn the instrument back
on.
If the problem occurs repeatedly,
contact Technical Support.
Error symptom
Possible cause
Corrective action
Summary of Contents for Typhoon 9200
Page 2: ...GE Healthcare Typhoon Operating Instructions Original instructions...
Page 3: ......
Page 13: ...1 Introduction 1 4 Control software 12 Typhoon Operating Instructions 28 9623 56 AA...
Page 27: ...3 Installation 3 5 Spare parts and accessories 26 Typhoon Operating Instructions 28 9623 56 AA...
Page 75: ...Appendix A Accessories A 4 Beamsplitters 74 Typhoon Operating Instructions 28 9623 56 AA...
Page 76: ......