March 2004
2-12
Programming Site Data
Entering Fluid Data
The
Fluid
submenu allows you to specify the fluid you are measuring,
as well as the Reynolds Correction factor and tracking windows.
While following the programming instructions, refer to Figure A-1 on
page A-1 of Appendix A,
Menu Maps
. Remember to record all
programmed data in Appendix C,
Data Records.
Note:
If you are in the PROG menu, proceed directly to Step 4. If
you scroll to a different option, press [ENT] twice to select
that option (once to enter and again to confirm the selection).
To access the
Fluid
submenu:
1.
Press [ESC]. The UTX878 enters the User Program.
2.
Press the [
X
] key until
PROG
is bracketed in the top left corner
and press [ENT].
3.
Use the [
S
] and [
T
] keys to scroll to the desired
Channel
and
press [ENT].
4.
Scroll to the
Fluid
submenu and press [ENT].
Entering Fluid Type
5.
Two options appear,
Fluid Type
and
Reynolds
. Scroll to
Fluid
Type
and press [ENT].
6.
Two other options appear,
Normal
and
Tracking
. Tracking refers
to Tracking Windows, which are used to detect the receive signal
when you are unsure of the fluid sound speed, or when the fluid
sound speed changes drastically under process conditions. Scroll
to the desired option and press [ENT].
7.
You can now select between
Preprogrammed
and
Other
fluids.
Scroll to the desired option, and press [ENT].
•
For Normal fluids, you can program the expected fluid
Temperature
. Scroll to the
Temperature
option and press
[ENT]. Then use the arrow keys to enter the process
temperature, and press [ENT].
•
For fluids monitored with a Tracking Window, the meter offers
the selections of
Water
(up to either 100 or 260°C) or
Oil
.
Scroll to the desired listing and press [ENT].
Summary of Contents for UTX878 Ultrasonic
Page 2: ...iii March 2004 The UTX878 Flow Transmitter and Transducers ...
Page 8: ...Chapter 1 ...
Page 20: ...March 2004 Installation 1 11 Figure 1 5 UTX878 Outline and Installation Dwg 712 1196 rev 2 ...
Page 22: ...March 2004 Installation 1 13 Figure 1 7 UTX878 Wiring Diagram Dwg 702 528 rev 2 ...
Page 23: ...Chapter 2 ...
Page 45: ...Chapter 3 ...
Page 52: ...Chapter 4 ...
Page 60: ...Chapter 5 ...
Page 65: ...Chapter 6 ...
Page 75: ...Chapter 7 ...
Page 80: ...Appendix A ...
Page 85: ...Appendix B ...
Page 86: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 ...
Page 88: ...Appendix C ...
Page 89: ...Data Records Site Data C 1 ...
Page 93: ...Appendix D ...
Page 94: ...Service Record Introduction D 1 Data Entry D 1 Diagnostic Parameters D 3 ...