Displaying Data
3-5
March 2004
Setting Screen
Contrast
For viewer convenience, you can reset the contrast level of the
display screen. To change the contrast:
1.
Press [ESC]. The UTX878 enters the
User Program
.
2.
Press the [
X
] key until
DISP
is bracketed in the upper left corner.
3.
The menu highlights the
Contrast
option. Press [ENT].
4.
Use the [
S
] and [
T
] keys to enter the desired contrast level from
0 to 100%. (The default value is 30%.) As you scroll, the screen
visibly changes contrast. When you have reached the desired
contrast level, press [ENT].
Note:
The display is visible over the entire UTX878 temperature
range without contrast adjustment.
5.
Press [ESC] to return to the
User Program
, or continue pressing
[ESC] to return to the display screen.
Setting the Number of
Screen Views
You can also choose to display either one or two measurement
parameters, on the same or different channels. To select the number
of views:
1.
Press [ESC]. The UTX878 enters the
User Program
.
2.
Press the [
X
] key until
DISP
is bracketed.
3.
Scroll to the
Views
option, and press [ENT].
4.
Scroll to the desired number of views, and press [ENT].
5.
Press [ESC] to return to the
User Program,
or continue pressing
[ESC] to return to the display screen.
Summary of Contents for UTX878 Ultrasonic
Page 2: ...iii March 2004 The UTX878 Flow Transmitter and Transducers ...
Page 8: ...Chapter 1 ...
Page 20: ...March 2004 Installation 1 11 Figure 1 5 UTX878 Outline and Installation Dwg 712 1196 rev 2 ...
Page 22: ...March 2004 Installation 1 13 Figure 1 7 UTX878 Wiring Diagram Dwg 702 528 rev 2 ...
Page 23: ...Chapter 2 ...
Page 45: ...Chapter 3 ...
Page 52: ...Chapter 4 ...
Page 60: ...Chapter 5 ...
Page 65: ...Chapter 6 ...
Page 75: ...Chapter 7 ...
Page 80: ...Appendix A ...
Page 85: ...Appendix B ...
Page 86: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 ...
Page 88: ...Appendix C ...
Page 89: ...Data Records Site Data C 1 ...
Page 93: ...Appendix D ...
Page 94: ...Service Record Introduction D 1 Data Entry D 1 Diagnostic Parameters D 3 ...