background image

Instructions 80-1265-18 AH

Vacugene

 

Vacuum Blotting Pump

GE Healthcare

Unpacking

Unpack the units and accessories. Do not discard
any packing materials until each item has been
checked against the packing list supplied with the
instrument.

Inspect all items for transit damage. If damage has
occurred, contact your local GE Healthcare
office and advise the transport company concerned.

Installing and operating the pump

1

Place the VacuGene™ Pump within reach of an electrical outlet.

Note:

This instrument is for indoor use only.

2

Check that the mains power switch on sawbuck panel of the 
pump is OFF (in the 0 position).

3

The mains power inlet, fuses and mains ON/OFF switch are 
located on the back panel of the vacuum pump.

4

Fit the mains cable into the mains power inlet and connect it to 
the mains power supply. Ensure that the mains cable is plugged 
into a fully grounded mains outlet.

5

The power inlet and power supply shall always be easily 
accessible.

6

Ensure that all air inlets and outlets are free from obstruction.

7

Make sure the instrument is placed on a stable laboratory work 
bench.

8

Using the tubing supplied with VacuGene XL Blotting Unit, 
connect the pump inlet on the front panel to the connection on 
the blotting unit. The needle valve controls the vacuum pressure 
and the vacuum gauge gives readings in the range 0-100 mbar 
with an accuracy of + 5%.

9

The low pressure VacuGene Vacuum Blotting Pump requires no 

regular maintenance.

To protect the pump, the liquid trap supplied with the VacuGene XL 
Blotting Unit should be installed between the pump and a blotting
unit. A “T” connector is provided to attach a second blotting unit to the 
vacuum pump, if required.

 .

Fig.1

Pump unit front panel.

Fig.2

Pump unit back panel.

WARNING! 

All repairs should be done by personnel 

authorized by GE Healthcare. Do not open any covers 
or replace any parts unless specifically stated in the 
instructions.

Vacuum

mbar

Reviews: