General Requirements
Venue 50 – Basic Service Manual
2-7
5447566-100 English Rev. 7
Site circuit breaker
It is recommended that the branch circuit breaker for the
Ultrasound system be readily accessible.
Site power outlets
A dedicated AC power outlet must be within reach of the
Ultrasound system without extension cords. Other outlets
adequate for the external peripherals, medical and test
equipment needed to support this Ultrasound system must also
be present within 1 m (3.2 ft.) of the Ultrasound system.
Electrical installation must meet all current local, state, and
national electrical codes.
Unit power plug
If the Ultrasound system arrives without a power plug, or with
the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.
Power stability requirements
Voltage drop-out
:
Max 10 ms.
Power transients
(all applications):
Less than 25% of nominal peak voltage for less than 1
millisecond for any type of transient, including line frequency,
synchronous, asynchronous, or aperiodic transients.
CAUTION
Power outage may occur. The Venue 50 requires a dedicated
single branch circuit. To avoid circuit overload and possible loss
of critical care equipment, make sure you do not have any
other equipment operating on the same circuit.
Summary of Contents for Venue 50
Page 5: ...Venue 50 Basic Service Manual i 3 5447566 100 English Rev 7...
Page 6: ...i 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Page 7: ...Venue 50 Basic Service Manual i 5 5447566 100 English Rev 7...
Page 8: ...i 6 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Page 9: ...Venue 50 Basic Service Manual i 7 5447566 100 English Rev 7...
Page 10: ...i 8 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Page 11: ...Venue 50 Basic Service Manual i 9 5447566 100 English Rev 7...
Page 12: ...i 10 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Page 22: ...i 20 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Page 52: ...Introduction 1 30 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Page 72: ...Site Preparations 2 20 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Page 160: ...Components and Functions Theory 5 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Page 204: ...Troubleshooting 7 38 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Page 228: ...Replacement Procedures 8 24 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Page 284: ...Care and Maintenance 10 36 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Page 288: ...Index 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Page 289: ......