8-34
Section 8-4 - System Modules- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-4-4
SSD Module Replacement Procedure
8-4-4-1
Tools
Phillips screwdriver
8-4-4-2
Time Required
15 min
8-4-4-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-4-4
SSD Module Removal Procedure
1) Remove all accessories.
2) Remove the following covers: RS Probe Cover, Lower Front eTower, Left side eTower, Right Side
eTower.
3.) Open the two fastening latches on the left side of the Front End door assembly and loosen the
captive screw.Then, open the Front End door.
4.) Fix the FE door with door support.
5.) Release the SSD Module captive screw and extract the SSD Module.
FRU Part #
Refer to
.
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
Figure 8-21 Removing the SSD Module B
..
Captive Screw
Summary of Contents for Venue R1
Page 477: ......