Chapter 8 - Replacement Procedures
8-43
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-4-7
Battery Module Replacement Procedure
8-4-7-1
Tools
Appropriate Phillips screwdriver.
8-4-7-2
Time Required
15 min
8-4-7-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-7-4
Battery Module Removal Procedure
1) Remove all accessories.
2) Remove the following covers: RS Probe Cover, Lower Front eTower cover, Right Side eTower
Cover, Left Side eTower Cover and MPB Door Cover:.
3.) Release the two captive screws securing each battery pack.
4.) Pull out the top and bottom battery packs.
FRU Part #
Refer to
.
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
MPB Door Cover Removal Procedure
Figure 8-30 Removing the Battery Pack
Battery Pack