Chapter 8 - Replacement Procedures
8-45
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-4-8
Back End (BE) Module Replacement Procedure
8-4-8-1
Tools
Appropriate Phillips screwdriver.
8-4-8-2
Time Required
15 min
8-4-8-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-8-4
BE Module Removal Procedure
1) Remove all accessories.
2) Remove the following covers: Lower Front eTower cover, Left Side eTower cover, and Right Side
eTower cover.
3) Remove the SSD module.
4) Remove MPB Front Metal Door.