Chapter 8 - Replacement Procedures
8-69
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-5-2
T-CFE Release Arm Replacement Procedure
8-5-2-1
Tools
Appropriate Phillips screwdriver.
8-5-2-2
Time Required
15 minutes
8-5-2-3
Preparation
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-5-2-4
CFE Board Release Arm Removal Procedure
The lever-shaped CFE Release Arm is located on the PSB Board.
1.) Remove all accessories.
2.) Remove the following covers: RS Probe Cover, Lower Front eTower, Left side eTower, and Right
Side eTower.
3) Remove the T-CFE Module:
4.) Unscrew the two lever holding screws and remove the CFE Release Arm (Lever) from the PSB
Board.
Figure 8-56 Location of CFE Release Arm
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
T-CFE Module Removal Procedure
CFE Release Arm
Summary of Contents for Venue R1