Section 8-6 - Mechanical Parts- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-6-2
MPB Guide L and Guide R Replacement Procedure
NOTE:
The procedure below is the same whether replacing MPB Guide L or MPB Guide R that are mirrored items
attached to the Receptacle Box side walls.
8-6-2-1
Tools
Appropriate Phillips screwdriver.
8-6-2-2
Time Required
15 min
8-6-2-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-6-2-4
MPB Guide L/R Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: RS Probe cover, Lower Front eTower, Side Covers, MPB Power Door
3.) Remove the MPB Module.
4.) Unscrew the two mounting screws that hold the Guide L/R.
FRU Part #
Refer to
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
MPB Door Cover Removal Procedure
•