Chapter 8 - Replacement Procedures
8-91
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-6-5
Riser Replacement Procedure
8-6-5-1
Tools
•
Allen key 2.5 mm.
•
LOCTITE® 243™
•
Forged Steel Ring Pliers
8-6-5-2
Time Required
10 min
8-6-5-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-6-5-4
Riser Removal Procedure
1.) Remove all accessories.
2.) Remove cockpit (monitor).
3.) Remove the following covers: RS Probe cover, Lower Front eTower cover, Left Side eTower cover,
Right Side eTower cover, Riser Thermal Cover and Upper Front eTower Cover.
4.) Remove scanner arm.
5.) Loosen and remove four screws securing the riser cover to the riser.
FRU Part #
Refer to
•
Accessories - Replacement Procedures
•
Cockpit (Monitor) Removal Procedure
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Upper eTower Front Cover Removal Procedure
•
Riser Thermal Cover Removal Procedure
•